Rapid detection of Escherichia coli by flow injection analysis coupled with amperometric method using an IrO2–Pd chemically modified electrode
详细信息    查看全文
文摘
An amperometric method for the rapid detection of Escherichia coli (E. coli) by flow injection analysis (FIA) using an IrO2–Pd chemically modified electrode (CME) was developed in this paper. The method is based on a good marker β-d-galactosidase which was found in E. coli strains. β-d-galactosidase was produced by the induction of isopropyl β-d-thiogalactopyranoside (IPTG) and released from E. coli cells through the permeabilization of both polymyxin B nonapeptide and lysozyme to E. coli cells wall. The released β-d-galactosidase could catalyze the hydrolysis of the substrate p-aminophenyl β-d-galactopyranoside (PAPG) in the culture medium to produce 4-aminophenol which was proportional to the concentration of E. coli. Hence, E. coli could be detected by the determination of 4-aminophenol. An IrO2–Pd CME, which showed high sensitivity in determination of 4-aminophenol, was prepared as the electro-detector in FIA. The amplified response current of 4-aminophenol obtained at the IrO2–Pd CME was linear with the concentration of E. coli ranging from 2.0 × 102 to 1.0 × 106 cfu/mL, the detection limit of this method to E. coli was 150 cfu/mL and the complete assay could be performed in 3 h.