A new PSD method was developed to suppress Compton-continuum in BEGe detectors by offline digital PSA. Pulse shapes from the preamplifier were digitized and saved while no other analog electronics were used. A decision parameter defined as the ratio of a smaller rise time parameter to r90 was presented and proved to be effective. The 60Co P/C ratio and the Cs-P/Co-C ratio can be improved by more than two and three times respectively.