Fully-integrated CMOS NMR receiver with co-integrated planar detection coil.
Spin sensitivity of 16001233&_mathId=si20.gif&_user=111111111&_pii=S1090780716001233&_rdoc=1&_issn=10907807&md5=c2365e235bc1ab19736783068976c4a6" title="Click to view the MathML source">3×101316001233&_mathId=si21.gif&_user=111111111&_pii=S1090780716001233&_rdoc=1&_issn=10907807&md5=1af9a675a19d0700ea1ad53bca09d5e5" title="Click to view the MathML source">1H spins/√Hz at 7 T and 340 pV/√Hz input-referred noise.
High-resolution sensitivity corrected MR microscopy images down to 5 μm resolution.
Sensitivity correction algorithm for planar receive coils in constant time imaging.
Derivation and validation of the relation between spin sensitivity and image SNR.