文摘
A national campaign was carried out in Italy for the application of coincidence-summing correction in γ-ray spectrometry. Twelve laboratories, belonging to the National Environmental Radioactivity Surveillance Network, took part in the campaign. They are equipped with γ-ray spectrometry systems based on p- and n-type HPGe detectors with different relative efficiencies. A simplified procedure was used for coincidence-summing correction. This procedure requires a single-photon, single-nuclide source to measure the peak-to-total efficiency ratio at just one photon energy value.