Quality of determinations obtained from laboratory reference samples used in the calibration of X-ray electron probe microanalysis of silicate minerals
详细信息    查看全文
文摘
Nine simple minerals and oxides, traditionally used as laboratory reference samples in the electron probe microanalysis (EPMA) of silicate minerals, have been quantitatively evaluated. Three separate series of data, comprising the average concentration, standard deviation, relative standard deviation, confidence interval and the z-score of data quality, were calculated for 21 control samples derived from calibrations obtained from three sets of reference samples: (1) simple minerals; (2) oxides; and (3) certified glass reference materials. No systematic difference was observed between the concentrations obtained from these three calibration sets when analyzed results were compared to certified compositions. The relative standard deviations obtained for each element were smaller than target values for all determinations. The z-score values for all elements determined fell within acceptable limits (−2<z<2) for concentrations ranging from 0.1 to 100 % . These experiments show that the quality of data obtained from laboratory reference calibration samples is not inferior to that from certified reference glasses. The quality of results obtained corresponds to the ‘applied geochemistry’ type of analysis (category 2) as defined in the GeoPT proficiency testing program. Therefore, the laboratory reference samples can be used for calibrating EPMA techniques in the analysis of silicate minerals and for controlling the quality of results.