Continuous supervision of extracorporeal O2 and CO2 gas transfer rates is proposed.
Only non-invasive sensors (e.g. in gas phase) are used (no blood gas analysis).
An extended Kalman filter for estimating O2 transfer rate is proposed.
Accuracy and performance were evaluated using in silico and in vivo data.
Results suggest that application in closed-loop control is feasible.