Twenty central incisors were endodontically treated with retrograde obturation by using amalgam, Biodentine, MTA, and Super-EBA (5 of each). They were placed in a skull with soft tissue simulation and scanned by using the Planmeca Promax Max with different kilovolt peaks (kVp): 66, 76, 84, and 96 with and without the use of metal artifact reduction (MAR) algorithm and with low, normal, and high resolution and high definition. The Dose Area Product was registered, and the effective dose was calculated.
Amalgam generated the highest amount of artifacts, whereas MAR and low resolution created fewer artifacts than other settings. The artifacts were also reduced with 96 kVp. The effective dose calculated with low resolution was remarkably lower than other resolutions.
When used as root-end filling material, Biodentine, MTA, and Super-EBA generated fewer artifacts than amalgam. The use of 96 kVp with MAR and low resolution also reduced artifacts on the image and at the same time generated the lowest effective dose.