Intraocular pressure measurement after DSAEK by iCare, Goldmann applanation and dynamic contour tonometry: A comparative study
详细信息    查看全文
文摘
Corneal thickness inevitably increases following Descemet's stripping automated endothelial keratoplasty (DSAEK), owing to the addition of a donor graft. The current study compares different devices in assessing post-DSAEK intraocular pressure (IOP).MethodsWe compared IOP values measured by the Goldmann tonometry (GAT), iCare rebound tonometry (iCare) and Pascal dynamic contour tonometry (PDCT) in eyes following DSAEK. Agreement between measurements was calculated with correlation analysis and Bland-Altman plots. Effects of keratometry, central, thickness (CCT), endothelial cell density (ECD) and axial length on IOP measurements were assessed with Pearson's correlation.ResultsTwenty eyes of 20 patients (mean age 74.3 ± 14.4, 14 females) post-DSAEK were included in this study. There was a high concordance between the IOP readings obtained by the three devices: a strong and significant correlation was found between GAT and PDCT (r = 0.94, P < 0.001) GAT and iCare (r = 0.86, P < 0.001) and iCare with PDCT (r = 0.81, P < 0.001). However, the iCare measurements were significantly and consistently lower than that obtained with GAT (ΔIOP = 1.68 ± 2.0, P = 0.002, 95% CI: 0.7&ndash;2.6) and with PDCT (ΔIOP = 1.61 ± 2.5, P = 0.01, 95% CI: 0.4&ndash;2.8). CCT, ECD, CCT, AXL, corneal curvature or astigmatism did not influence IOP measurement by any instrument.ConclusionsIOP measurement with three different techniques (applanation, rebound and dynamic contour) showed good correlations, despite an increased corneal thickness following DSAEK. However, the iCare, which is based on a rebound tonometry showed significant lower IOP then the two other methods. This should be taken into account when evaluating patients post DSAEK.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700