FD-OCT and IVUS images sequentially obtained from a phantom coronary model and 57 stented human coronary arteries were compared between each other and with QCA.
Lumen area (LA) by IVUS was 10.1% larger (6.43 ± 0.09 mm2) while by FD-OCT was similar (5.78 ± 0.09 mm2) to actual phantom LA (5.72 mm2); IVUS vs. FD-OCT stent area (SA) was 4.2% larger. In human coronary artery, diameter by QCA was smaller than by IVUS and OCT in reference (by 10.5% and 3.5%, both p < 0.001) and stented (3.6%, p < 0.001; and 1.7%, p = 0.012) segments. IVUS vs. FD-OCT distal reference LA was significantly larger (6.19 ± 2.18 mm2 vs. 5.49 ± 2.49 mm2, p < 0.001, respectively), and SA was numerically larger (7.42 ± 2.28 mm2 vs. 7.22 ± 2.48 mm2, p = 0.059) with larger discrepancy in reference (11.3%) than stented (2.7%) segments. IVUS vs. FD-OCT correlation for diameter was significantly higher for stented than reference segments (R2 = 0.8670 vs. 0.7351, p = 0.047), while numerically higher for area (R2 = 0.8663 vs. 0.7806, p = 0.157).
In phantom model and human coronary arteries, IVUS vs. FD-OCT measurements were larger, particularly in non-stented than stented segments, and diameter was smaller by QCA vs. IVUS or FD-OCT. Despite undefined clinical significance, said discrepancy warrants consideration.