Three types of two-piece implant CAD models were designed: external joint with a conical tapered neck (EJ), internal joint with a straight neck (IJ), and conical joint with a reverse conical neck (CJ). 3D FEA was performed with the setting of a ¡°contact¡± condition at the component interface, and stress distribution in the peri-implant bone and abutment micromovement were analyzed.
The shear stress was concentrated on the mesiodistal side of the cortical bone for EJ. EJ had the largest amount of abutment micromovement. While the von Mises and shear stresses around the implant neck were concentrated on the labial bone for IJ, they were distributed on the mesiodistal side of the cortical bone for CJ. CJ had the least amount of abutment micromovement.
Implants with a conical joint with an abutment and reverse conical neck design may effectively control occlusal overloading on the labial bone and abutment micromovement.