Addressing a systematic bias in carbon dioxide flux measurements with the EC150 and the IRGASON open-path gas analyzers
详细信息    查看全文
文摘

CO2 flux measurement errors (IRGASON & EC150) scale with kinematic temperature flux.

Relative errors are most pronounced when true CO2 flux is small and heat flux large.

Using a fast-response air temperature to scale CO2 absorption minimizes bias.

Agreement between open- and closed-path IRGA CO2 fluxes is substantially improved.

Fast-instead of slow-response air temperature should be used to scale CO2 absorption.