We evaluated the dosimetric uncertainty of the reflective-type radiochromic film.
The sharp increase in the pixel value occurred within the 3 h after irradiation.
The uncertainty of the stochastic noise was independent of elapsed time.
The uncertainty of the density growth was decreased within 1 h after irradiation.
Total uncertainty decreased 1 h after irradiation and remained constant afterward.