Sensitive square wave anodic stripping voltammetric determination of Cd2+ and Pb2+ ions at Bi/Nafion/overoxidized 2-mercaptoethanesulfonate-tethered polypyrrole/glassy carbon electrode
详细信息    查看全文
文摘
Square wave anodic stripping voltammetry was applied to detect Pb2+ and Cd2+ ions at a glassy carbon electrode modified with overoxidized 2-mercaptoethanesulfonate (MES)-tethered polypyrrole, Nafion and Bi. The PPy-MES interaction was examined by quartz crystal microbalance. Under optimum conditions (5-min preconcentration at 鈭?.2 V vs. SCE in pH 4.0 acetate buffer containing 300 渭g L鈭? Bi3+), the linear calibration curves ranged from 0.05 to 35 渭g L鈭? for Pb2+ and 0.1 to 25 渭g L鈭? for Cd2+, respectively, and the limits of detection (S/N = 3, S is the signal and N is the noise level) were 0.03 渭g L鈭? for Pb2+ and 0.04 渭g L鈭? for Cd2+. This method was successfully applied for Pb2+ and Cd2+ analysis in real water samples.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700