Biostability of micro-photodiode arrays for subretinal implantation
详细信息    查看全文
文摘
Micro-photodiode arrays based on semiconductor chip technology are being developed to replace degenerated photoreceptor cells in the retina. Electric current is generated in tiny micro-photodiodes and delivered to the adjacent tissue by micro-electrodes. One of the main requirements of a sub-retinal implantable device is long-term stability versus corrosion in vivo (biostability). Biostability of micro-photodiode arrays (MPDA) was investigated in vitro and in vivo. No significant damage was found on chips immersed for up to 21 months in saline solution. Under in vivo conditions, however, the silicon oxide passivation layer of the chip was dissolved within a period of about 6–12 months. Subsequently, the underlying silicon was corroded. In contrast, stimulation electrodes consisting of titanium nitride were well preserved both in vitro and in vivo. The deterioration of the electrical properties of the micro-photodiodes correlated with the morphological damage observed. Strategies aiming at the development of an improved biostable encapsulation of neurotechnological implants have to be investigated and will be discussed briefly.