In Vitro Evaluation of Focused Ultrasound-Enhanced TNK-Tissue Plasminogen Activator-Mediated Thrombolysis
详细信息    查看全文
文摘
The low and incomplete recanalization performance of thrombolytic therapy in patients with acute ischemic stroke has created the need to use focused ultrasound (FUS) energy as a way to enhance thrombolysis efficiency (sonothrombolysis). Using an in vitro flow model, the role of various parameters involved in FUS-enhanced tenecteplase (TNK-tPA [tissue plasminogen activator])-mediated thrombolysis was evaluated.

Materials and methods

Fully retracted porcine blood clots were used for the proposed parametric studies. A spherically FUS transducer (4 cm diameter), focusing at 10 cm and operating at 1 MHz, was used. Pulsed ultrasound protocols were applied that maintained temperature elevation at the focus that never exceeded 1°C. Thrombolysis efficiency was measured as the relative reduction in the mass of the clot.

Results

The role of various properties on thrombolysis efficacy was examined. These various properties are the acoustic power, the TNK-tPA concentration, the flow rate, the exposure time, the pulse length, the pulse repetition frequency, the duty factor, the formation of standing waves, the acoustic medium, and the administration of microbubbles. Study results have demonstrated that the parameters examined influenced thrombolysis efficacy and the degree of thrombolysis achieved by each parameter was measured.

Conclusions

Study findings helped us to optimize the treatment protocol for 1 MHz pulsed FUS that maximizes the thrombolytic efficacy of TNK-tPA, which potentially could be applied for therapeutic purposes. The outcome of the study showed poor thrombolysis efficacy, as with 30 minutes of FUS treatment only 370 mg of clot was removed.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700