Flow rates through intravenous access devices: an in vitro study
详细信息    查看全文
文摘
Fluid administration using intravenous (IV) access devices is required in many settings. There are a lack of quantitative data comparing traditional cannulas and modern access devices. We aimed to investigate flow rates through modern intravenous access devices using an in vitro system.

Design

This is an experimental study.

Setting and measurements

Rates of flow of intravenous fluids (crystalloid and colloid) were measured through various access devices using a uroflowmeter. Standardized conditions and repeat measurements ensured validity. Fluid was administered with or without the addition of a pressure bag and needle-free valve.

Main results

Increasing the size of cannulas improved flow. Fourteen-gauge cannulas had significantly higher mean flow rates compared to 14G central venous lines in all conditions (136% higher with no pressure bag/valve; 95% CI, + 130% to + 152%; P < .001). Both the emergency infusion device and rapid infusion catheter produced significantly increased mean flows compared to a 14G cannula (12% higher for emergency infusion catheter; 95% CI, + 7% to + 15%; P = .008, and 15% higher for rapid infusion catheter; 95% CI, + 12% to + 21%; P = .004). The needle-free valve significantly impaired flow on 16G and wider IV access devices (36% lower with no pressure bag using 14G cannula; 95% CI, − 29% to − 46%; P = .003), but flow reductions in narrower IV access were insignificant. Pressure bags significantly improved flow in all devices, in all combinations.

Conclusions

Flow rates in IV devices can be maximized by pressure bag use and removal of needle-free valves. The rapid infusion catheter and emergency infusion catheter allow some increase in flow over a 14G cannula. Familiarity with varying flow rates across IV access devices could better inform clinical decisions.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700