Microfabricated Gas Chromatograph for Rapid, Trace-Level Determinations of Gas-Phase Explosive Marker Compounds
详细信息    查看全文
文摘
A prototype microfabricated gas chromatograph (渭GC) adapted specifically for the rapid determination of selected gas-phase marker compounds of the explosive 2,4,6-trinitrotoluene (TNT) at sub-parts-per-billion (<ppb) concentrations in complex mixtures is described. Si-microfabricated focuser, separation column, and sensor array components are integrated with a high-volume sampler of conventional construction to reduce analysis time and the limit of detection (LOD). The primary markers selected as target analytes were 2,4-dinitrotoluene (2,4-DNT, a persistent impurity of TNT) and 2,3-dimethyl-2,3-dinitrobutane (DMNB, a taggant), along with 2,6-dinitrotoluene (2,6-DNT, a less-prominent TNT impurity), which was also included in numerous tests. Selective preconcentration, on-column focusing, temperature-programmed chromatographic separation, and sensor array detection/recognition facilitated determinations of the primary markers in the presence of 20 (or more) interferences within 2 min under laptop control. Estimated LODs of 2.2, 0.48, and 0.86 ng were achieved for DMNB, 2,6-DNT, and 2,4-DNT, respectively, which correspond to 0.30, 0.067, and 0.12 ppb in each 1-L air sample collected.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700