Rapid and Simultaneous Determination of Essential Minerals and Trace Elements in Human Milk by Improved Flame Atomic Absorption Spectroscopy (FAAS) with Microwave Digestion
详细信息    查看全文
文摘
A method for the simultaneous and economical determination of many trace elements in human milk is developed. Two multi-element hollow cathode lamps (HCLs) were used instead of single-element HCLs to improve the sample throughput of flame atomic absorption spectroscopy (FAAS). The microwave digestion of milk is optimized prior to detection, and the performance characteristics of the improved analysis method are identified. Clinical samples are detected by both FAAS and inductively coupled plasma−optical emission spectroscopy (ICP−OES) for methodology evaluation. Results reveal that the proposed FAAS with multi-element HCLs could determine six essential minerals and trace elements within 15 min. This method provides a linear analytical range of 0.01−10 mg L−1. For Ca, Cu, Fe, Mg, Mn, and Zn, the limits of determination are 1.5, 3, 1.8, 2.2, 2.1, and 1.3 μg L−1, respectively. The mean relative standard deviations (RSDs) of intra- and interassays are lower than 7%. Excellent operational characteristics of rapidity, simplicity, and economy make the proposed method a promising one for the quantification of trace elements in human milk in clinics of underdeveloped areas.

Keywords:

Atomic absorption spectroscopy (AAS); human milk; microwave digestion; multi-element; trace elements

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700