A competitive direct enzyme-linked immunosorbent assay (CD-ELISA) for histamine in cheese wascompared with a reversed-phase liquid chromatography (RP-HPLC) method. Cheese was homogenized with phosphate-buffered saline (PBS), centrifuged, and filtered, and the supernatant wasdiluted with PBS for CD-ELISA. For RP-HPLC, biogenic amines (histamine, tyramine, putrescine,and cadaverine) were derivatized with 9-fluorenylmethylchloroformate, followed by reversed-phasechromatography and fluorescence detection. Detection limits and mean recoveries (10-1000 mg/kg) were 2 mg/kg and 93% for CD-ELISA and 1 mg/kg and 99% for RP-HPLC, respectively. Analysisof 50 commercial cheeses according to both methods showed good agreement for histamine (r =0.979; concentration range = 2-1800 mg/kg). At a threshold level of 10 mg/kg, the ELISA gave nofalse-negative and three false-positive results. The results show that the ELISA is suitable for thedetermination of histamine in cheese.Keywords: Biogenic amines; histamine; cheese; enzyme immunoassay; liquid chromatography