文摘
A prototype is introduced based on the transversal modulation ion mobility spectrometry (TMIMS) technique, which provides a continuous output of mobility-selected ions, greatly easing the synchronization between different analyzing stages. In the new architecture, two stages of filtration are used to drastically reduce the background produced by one stage alone. Two-stages TMIMS was coupled with two different atmospheric pressure interface mass spectrometers (MS). The new system enables IMS-IMS-MS analysis and other modes of operation: IMS prefiltration, IMS-IMS, and full transmission mode. It provides a resolving power R > 60 in IMS mode, and R > 40 in each stage of IMS-IMS mode. 2-Propanol vapors were introduced in one of the stages to enhance the mobility variations, and their effect was studied on a set of tetraalkylammonium ions. We found that concentrations as low as 1% (in partial pressure) produce mobility variations as high as 20%, which suggest that IMS-IMS separation using dried N2 (in one stage) and a dopant (in the other stage), could be a very powerful way to enhance the separation capacity of the IMS-IMS prefiltration approach.