Demonstration of a Mid-Infrared Cavity Enhanced Absorption Spectrometer for Breath Acetone Detection
详细信息    查看全文
文摘
A high-resolution absorption spectrum of gaseous acetone near 8.2 渭m has been taken using both Fourier transform and quantum cascade laser (QCL)-based infrared spectrometers. Absolute absorption cross sections within the 1215鈥?222 cm鈥? range have been determined, and the spectral window around 1216.5 cm鈥? (蟽 = 3.4 脳 10鈥?9 cm2 molecule鈥?) has been chosen for monitoring trace acetone in exhaled breath. Acetone at sub parts-per-million (ppm) levels has been measured in a breath sample with a precision of 0.17 ppm (1蟽) by utilizing a cavity enhanced absorption spectrometer constructed from the QCL source and a linear, low-volume, optical cavity. The use of a water vapor trap ensured the accuracy of the results, which have been corroborated by mass spectrometric measurements.