Method for Characterizing Nanoscale Wear of Atomic Force Microscope Tips
详细信息    查看全文
文摘
Atomic force microscopy (AFM) is a powerful tool for studying tribology (adhesion, friction, and lubrication) at the nanoscale and is emerging as a critical tool for nanomanufacturing. However, nanoscale wear is a key limitation of conventional AFM probes that are made of silicon and silicon nitride (SiNx). Here we present a method for systematically quantifying tip wear, which consists of sequential contact-mode AFM scans on ultrananocrystalline diamond surfaces with intermittent measurements of the tip properties using blind reconstruction, adhesion force measurements, and transmission electron microscopy (TEM). We demonstrate direct measurement of volume loss over the wear test and agreement between blind reconstruction and TEM imaging. The geometries of various types of tips were monitored over a scanning distance of approximately 100 mm. The results show multiple failure mechanisms for different materials, including nanoscale fracture of a monolithic Si tip upon initial engagement with the surface, film failure of a SiNx-coated Si tip, and gradual, progressive wear of monolithic SiNx tips consistent with atom-by-atom attrition. Overall, the method provides a quantitative and systematic process for examining tip degradation and nanoscale wear, and the experimental results illustrate the multiple mechanisms that may lead to tip failure.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700