文摘
A semiautomated device for solubility measurements in solution volumes as low as 1 mL wasdeveloped for pharmaceutical applications. The device operates nonisothermally, and can measuremultiple samples simultaneously. Solubilities of model compounds such as paracetamol, mannitol,adipic acid, and glycine were measured and showed a maximum deviation of 5 wt % comparedto literature data. Extrapolation of solubility data in the form of activity coefficients was foundin good agreement with the literature value over a wide range of temperatures for paracetamolin 2-propanol and adipic acid in ethanol, while some deviation was observed for mannitol inwater. The use of the solubility-measuring device to perform solvent screening on multiplesolvents for early-stage pharmaceutical compounds was also examined. Results indicate thatthe slope of the solubility curve with temperature can be estimated with limited data for multiplesolvents.