SFE-GC with Quantitative Transfer of the Extraction Effluent to a Megabore Capillary Column
详细信息    查看全文
  • 作者:Mark A. Stone and Larry T. Taylor
  • 刊名:Analytical Chemistry
  • 出版年:2000
  • 出版时间:July 15, 2000
  • 年:2000
  • 卷:72
  • 期:14
  • 页码:3085 - 3092
  • 全文大小:118K
  • 年卷期:v.72,no.14(July 15, 2000)
  • ISSN:1520-6882
文摘
Supercritical fluid extraction was directly coupled with gaschromatography such that the entire extraction effluentwas directed onto the GC column. A sealed connectionbetween the restrictor and the column ensured that noloss of analyte could occur. The elevated pressure generated in the column itself resulted in a reduced flow ratethrough the column, making efficient trapping possible.Metal-jacketed capillary columns were used as they cantolerate the elevated pressures required. In addition tomaximizing sensitivity, very practical extraction parameters were possible. For a 25-min extraction, with a 2.2mL/min flow rate, and the column at 40 C, analytes asvolatile as C-14 were trapped and separated. Trappingefficiency improved further with shorter extraction times,mild cooling of the column, or lower flow rates.