Development and Biological Evaluation of Inkjet Printed Drug Coatings on Intravascular Stent
详细信息    查看全文
文摘
Inkjet-printing technology was used to apply biodegradable and biocompatible polymeric coatings of poly(d,l-lactide) with the antiproliferative drugs simvastatin (SMV) and paclitaxel (PCX) on coronary metal stents. A piezoelectric dispenser applied coating patterns of very fine droplets (300 pL) and inkjet printing was optimized to develop uniform, accurate and reproducible coatings of high yields on the stent strut. The drug loaded polymeric coatings were assed by scanning electron microscopy (SEM), atomic force microscopy (AFM), and transition thermal microscopy (TTM) where a phase separation was observed for SMV/PLA layers while PCX showed a uniform distribution within the polymer layers. Cytocompatibility studies of PLA coatings showed excellent cell adhesion with no decrease of cell viability and proliferation. In vivo stent implantation studies showed significant intrastent restenosis (ISR) for PCX/PLA and PLA plain coatings similar to marketed Presillion (bare metal) and Cypher (drug eluting) stents. The investigation of several cytokine levels after 7 days of stent deployment showed no inflammatory response and hence no in vivo cytotoxicity related to PLA coatings. Inkjet printing can be employed as a robust coating technology for the development of drug eluting stents compared to the current conventional approaches.