Rapid Quantification of Digitoxin and Its Metabolites Using Differential Ion Mobility Spectrometry-Tandem Mass Spectrometry
详细信息    查看全文
文摘
This study focuses on the quantitative analysis of the cardiac glycoside drug digitoxin and its three main metabolites digitoxigenin-bisdigitoxose, digitoxigenin-monodigitoxose, and digitoxigenin using electrospray ionization-differential ion mobility spectrometry-tandem mass spectrometry (ESI-DMS-MS/MS). Despite large molecular weight differences, gas-phase separation of the four compounds in the DMS drift cell was not possible, even by utilizing additional volatile chemical modifiers. Baseline separation was achieved after adduct formation with alkali metal ions, however, and efficiency was shown to improve with increasing size of the alkali ion, reaching optimum conditions for the largest cesium ion. Subsequently, an assay was developed for quantification of digitoxin and its metabolites from human serum samples and its analytical performance assessed in a series of proof-of-concept experiments. The method was applied to spiked human serum pools with concentration levels between 2 and 80 ng/mL. After a short reversed-phase chromatographic step for desalting the sample, rapid DMS separation of the analytes was carried out, resulting in a total run time of less than 1.5 min. The instrumental method showed good repeatability; the calculated coefficients of variation ranged from 2% to 13%.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700