Intraoperative Fracture of a Pyrocarbon PIP Total Joint—A Case Report
详细信息    查看全文
  • 作者:Martin Skie (1)
    Nicholas Gove (1)
    Despina Ciocanel (1)
  • 关键词:Proximal interphalangeal joint ; Proximal interphalangeal joint prosthesis ; Ascension PIP total joint
  • 刊名:HAND
  • 出版年:2007
  • 出版时间:September 2007
  • 年:2007
  • 卷:2
  • 期:3
  • 页码:90-93
  • 全文大小:139KB
  • 参考文献:1. Ascension PIP: Summary of safety and probable benefit (http://www.fda.gov/cdrh/pdf/h010005b.pdf).
    2. Beckenbaugh RD. Arthroplasty of the metacarpophalangeal joint using pyrocarbonate implants. Orthopade 2003;32(9):794-, Sep. [Article in German]. CrossRef
    3. Bravo C, Marco R, Hormel K, Bechenbaugh R. Pyrolytic carbon proximal interphalangeal joint arthroplasty: results with minimum two-year follow-up evaluation. J Hand Surg 2007;32A(1):1-1.
    4. Cook SD, Beckenbaugh RD, Redondo J, Popich LS, Klawitter JJ, Linscheid RL. Long-term follow-up of pyrolytic carbon metacarpophalangeal implants. J Bone Jt Surg Am 1999;81(5):635-8.
    5. Haubold AD. On the durability of pyrolytic carbon in vivo. Med Prog Technol 1994;20(3-):201-, Review.
    6. Hormel K, Beckenbaugh R. Preliminary report of total joint replacement of the PIP joint with a pyrolytic carbon implant. Orthopedics 2003;26(12 suppl):s1290. (abstract).
    7. Linscheid RL. Implant arthroplasty of the hand: retrospective and prospective considerations. J Hand Surg [Am] 2000;25(5):796-16, Sep. CrossRef
    8. Moutet F, Guinard D, Gerard P, De Soras X, Ranc R, Moreau C. A new titanium-carbon finger joint implant. Apropos of 15 initial cases. Ann Chir Main Memb Super 1994;13(5):345-3. [Article in French]. CrossRef
    9. Munshy P, Trail IA, Stanley JK, Rebelo N, Freire A, Mota da Costa J. Pyrocarbon PIP arthroplasty for post trauma painful joints. J Hand Surg [Am] 2003;28B(suppl):44. (abstract). CrossRef
    10. Ryder JK, Cao H. Structural integrity assessment of heart valve prostheses: a damage tolerance analysis of the CarboMedics Prosthetic Heart Valve. J Heart Valve Dis 1996;5(Suppl 1):S86-6, Jun, Review.
    11. Schulz M, Muller-Zimmermann A, Behrend M, Krimmer H. Early results of proximal interphalangeal joint replacement with pyrolytic carbon prosthesis (Ascension) in idiopathic and post-traumatic arthritis. Handchir Mikrochir Plast Chir 2005;37(1):26-4, Feb. [Article in German]. CrossRef
    12. Stutz N, Meier R, Krimmer H. Pyrocarbon prosthesis for finger interphalangeal joint replacement. Experience after one year. Unfallchirurg 2005;108(5):365-, May. [Article in German]. CrossRef
  • 作者单位:Martin Skie (1)
    Nicholas Gove (1)
    Despina Ciocanel (1)

    1. Department of Orthopaedic Surgery, University of Toledo, Health Science Campus, 3065 Arlington Ave, Dowling Hall, Toledo, OH, 43614, USA
文摘
Pain or dysfunction of the finger joints due to arthritis or traumatic injuries that fail medical management may necessitate arthroplasty or joint replacement. The goals of the finger joint implant arthroplasty are to relieve pain, to correct deformity, and to improve the function and appearance of the hand. Several prosthetic implants have been used for the replacement of the proximal interphalangeal (PIP) joint. Pyrocarbon materials, a form of pyrolytic carbon, a ceramic-like material, have proven to be strong, durable, resistant to wear and nonreactive in the body. The Ascension PIP pyrocarbon total joint is a bicondylar, anatomically shaped, articulating implant that allows joint flexion–extension, while providing some restriction of adduction–abduction motion. A review of the literature of pyrocarbon PIP prosthesis reveals little clinical data. The case of a 33-year-old man with posttraumatic arthritis proximal interphalangeal joint right long finger is reported. The case was treated surgically using the Ascension PIP total joint. During the insertion of the implant, the proximal component fractured at the sub-articular collar. The component was removed, and a new implant was inserted without complication. Critical evaluation revealed that there was an inadequate resection of the volar lip resulting in a stress riser on the implant during impaction. Careful attention to this technical point will hopefully minimize the occurrence of this complication as the availability and use of these implants increases.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700