The joint space difference index (JSDI) is a newly developed radiographic index which can quantitatively assess joint space narrowing progression of rheumatoid arthritis (RA) patients by using an image subtraction method on a computer. The aim of this study was to investigate the reliability of this method by non-experts utilizing RA image evaluation. Four non-experts assessed JSDI for radiographic images of 510 metacarpophalangeal joints from 51 RA patients twice with an interval of more than 2 weeks. Two rheumatologists and one radiologist as well as the four non-experts examined the joints by using the Sharp–van der Heijde Scoring (SHS) method. The radiologist and four non-experts repeated the scoring with an interval of more than 2 weeks. We calculated intra-/inter-observer reliability using the intra-class correlation coefficients (ICC) for JSDI and SHS scoring, respectively. The intra-/inter-observer reliabilities for the computer-based method were almost perfect (inter-observer ICC, 0.966–0.983; intra-observer ICC, 0.954–0.996). Contrary to this, intra-/inter-observer reliability for SHS by experts was moderate to almost perfect (inter-observer ICC, 0.556–0.849; intra-observer ICC, 0.589–0.839). The results suggest that our computer-based method has high reliability to detect finger joint space narrowing progression in RA patients.