Sensitive Determination of Bromate in Water Samples by Capillary Electrophoresis Coupled with Electromembrane Extraction
详细信息    查看全文
  • 作者:Xiaoli Zhang ; Lin Guo ; Dongxue Zhang ; Xinxin Ge ; Jiannong Ye…
  • 关键词:Bromate ; Drinking water ; Electromembrane extraction ; Capillary zone electrophoresis ; Capacitively contactless conductivity detection
  • 刊名:Food Analytical Methods
  • 出版年:2016
  • 出版时间:February 2016
  • 年:2016
  • 卷:9
  • 期:2
  • 页码:393-400
  • 全文大小:396 KB
  • 参考文献:Barron L, Paull B (2006) Simultaneous determination of trace oxyhalides and haloacetic acids using suppressed ion chromatography-electrospray mass spectrometry. Talanta 69:621–630CrossRef
    Bodor R, Kaniansky D, Masar M, Silleova K, Stanislawski B (2002) Determination of bromate in drinking water by zone electrophoresis-isotachophoresis on a column-coupling chip with conductivity detection. Electrophoresis 23:3630–3637CrossRef
    Chen JH, Wang KE, Jiang SJ (2007) Determination of iodine and bromine compounds in foodstuffs by CE-inductively coupled plasma MS. Electrophoresis 28:4227–4232CrossRef
    de Borba BM, Rohrer JS, Pohl CA, Saini C (2005) Determination of trace concentrations of bromate in municipal and bottled drinking waters using a hydroxide-selective column with ion chromatography. J Chromatogr A 1085:23–32CrossRef
    Eickhorst T, Seubert A (2004) Germanium dioxide as internal standard for simplified trace determination of bromate, bromide, iodate and iodide by on-line coupling ion chromatography-inductively coupled plasma mass spectrometry. J Chromatogr A 1050:103–109CrossRef
    EPA. Stage 1 disinfectants and disinfection byproducts rule (1998-12-16). http://​water.​epa.​gov/​lawsregs/​rulesregs/​sdwa/​stage1/​factsheet.​cfm . Accessed 23 Dec. 2013
    Fracassi da Silva JA, do Lago CL (1998) An oscillometric detector for capillary electrophoresis. Anal Chem 70:4339–4343CrossRef
    Fukushi K, Yamazaki R, Yamane T (2009) Determination of bromate in highly saline samples using CZE with on-line transient ITP. J Sep Sci 32:457–461CrossRef
    Gjelstad A, Pedersen-Bjergaard S (2011) Electromembrane extraction: a new technique for accelerating bioanalytical sample preparation. Bioanalysis 3:787–797CrossRef
    Gries T, Sitorius E, Giesecke A, Schlegel V (2008) Feasibility of using capillary zone electrophoresis with photometric detection for the trace level detection of bromate in drinking water. Food Addit Contam A 25:1318–1327CrossRef
    IARC (1990) Some naturally occurring and synthetic food components, furocoumaris and ultraviolet radiation: potassium bromate. Monographs on the Evaluation of the Carcinogenic Risk to Human. Vol. 40. Geneva
    Jackson PE (2001) Determination of inorganic ions in drinking water by ion chromatography. Trends Anal Chem 20:320–329CrossRef
    Kitamaki Y, Takeuchi T (2004) Cyclodextrin-aided determination of iodate and bromate in drinking water by microcolumn ion chromatography with precolumn enrichment. Anal Sci 20:1399–1402CrossRef
    Kuban P, Hauser PC (2009) Ten years of axial capacitively coupled contactless conductivity detection for CZE—a review. Electrophoresis 30:176–188CrossRef
    Kurokawa Y, Maekawa A, Takahasahi M (1990) Toxicity and carcinogenicity of potassium bromate—a new renal carcinogen. Environ Health Perspect 87:309–335
    Kvasnicka F, Janda V, Rousova D, Manda J, Kollerova L (2005) Determination of inorganic oxyhalides in drinking water by on-line coupled capillary isotachophoresis-capillary zone electrophoresis. Cent Eur J Chem 3:137–145
    Lawal W, Ghandi J, Zhang C (2010) Direct injection, simple and robust analysis of trace-level bromate and bromide in drinking water by ion chromatography with suppressed conductivity detection. J Chromatogr Sci 48:537–544CrossRef
    Li WL, Pan YL, Liu Y, Zhang XL, Ye JN, Chu QC (2014) Simultaneous determination of eight typical biogenic amines by CZE with capacitively coupled contactless conductivity detection. Chromatographia 77:287–292CrossRef
    Liu LB, Pang YH, Qian H, Xu HQ (2011) Determination of bromate in drinking water by capillary electrophoresis-electrochemical detection with MWCNTs/PPy/PMo12 modified electrode. J Anal Sci 27:479–482CrossRef
    Marak J, Stanova A, Vavakova V, Hrenakova M, Kaniansky D (2012) On-line capillary isotachophoresis-capillary zone electrophoresis analysis of bromate in drinking waters in an automated analyzer with coupled columns and photometric detection. J Chromatogr A 1267:252–258CrossRef
    Marothu VK, Gorrepati M, Vusa R (2013) Electromembrane extraction—a novel extraction technique for pharmaceutical, chemical, clinical and environmental analysis. J Chromatogr Sci 51:619–631CrossRef
    Mbuna J, Ngassapa FN, Motomizu S (2008) Use of weak ion association in the separation of inorganic anions by capillary electrophoresis with specific application to simultaneous-trace determination of bromate and iodate in drinking water. B Chem Soc Ethiopia 22:1–9
    Michalski R (2005) A comparison of direct and indirect ion chromatographic methods for the determination of bromide/bromate ions in drinking water. Chem Anal-Warsaw 50:583–591
    Michalski R, Lyko A (2013) Bromate determination: state of the art. Crit Rev Anal Chem 43:100–122CrossRef
    Ministry of Health of the People’s Republic of China (2006) Standards for drinking water quality (GB/T 5749-2006)
    Pan YL, Chen F, Zhang MY, Wang TQ, Xu ZC, Zhang W, Chu QC, Ye JN (2013) Sensitive determination of chloroanilines in water samples by hollow fiber-based liquid-phase microextraction prior to capillary electrophoresis with amperometric detection. Electrophoresis 34:1241–1248CrossRef
    Pang YH, Liu LB, Shen XF, Qian H (2012) Determination of bromate in drinking water by capillary electrophoresis coupled with chemically modified electrode electrochemical detection. Sci China (Series B: Chem) 42:157–163
    Pedersen-Bjergaard S, Rasmussen KE (2006) Electrokinetic migration across artificial liquid membranes: new concept for rapid sample preparation of biological fluids. J Chromatogr A 1109:183–190CrossRef
    Song R, Donohoe C, Minear R, Westerhoff P, Ozekin K, Amy G (1996) Empirical modelling of bromate formation during ozonation of bromide-containing waters. Water Res 30:1161–1168CrossRef
    Takayanag T, Ishida M, Mbuna J, Driouich R, Motomizu S (2006) Determination of bromate ion in drinking water by capillary zone electrophoresis with direct photometric detection. J Chromatogr A 1128:298–302CrossRef
    Wagner HP, Pepich BV, Hautman DP, Munch DJ (1999) Analysis of 500-ng/l levels of bromate in drinking water by direct-injection suppressed ion chromatography coupled with a single, pneumatically delivered post-column reagent. J Chromatogr A 850:119–129CrossRef
    Wang Y, Qiu J, Zhao GH (2009) Review on determination of bromate in drinking water. Food Sci (Beijing, China) 30:460–464
    WHO (2004) Guidelines for drinking-water quality. WHO, Geneva
    Zemann AJ, Schnell E, Volgger D, Bonn GK (1998) Contactless conductivity detection for capillary electrophoresis. Anal Chem 70:563–567CrossRef
    Zhang HJ, Gavina J, Feng YL (2011) Understanding mechanisms of pressure-assisted electrokinetic injection: application to analysis of bromate, arsenic and selenium species in drinking water by capillary electrophoresis-mass spectrometry. J Chromatogr A 1218:3095–3104CrossRef
  • 作者单位:Xiaoli Zhang (1)
    Lin Guo (1)
    Dongxue Zhang (1)
    Xinxin Ge (1)
    Jiannong Ye (1)
    Qingcui Chu (1)

    1. School of Chemistry and Molecular Engineering, East China Normal University, Shanghai, 200241, China
  • 刊物类别:Chemistry and Materials Science
  • 刊物主题:Chemistry
    Food Science
    Chemistry
    Microbiology
    Analytical Chemistry
  • 出版者:Springer New York
  • ISSN:1936-976X
文摘
Electromembrane extraction (EME) as a novel sample preparation technique was firstly applied for the purification and enrichment of bromate (BrO3 −) in drinking water prior to capillary zone electrophoresis with capacitively coupled contactless conductivity detection (CZE-C4D). BrO3 −, as the primary disinfection by-product of ozonation, could be well separated with the major inorganic anions coexisting in water samples using a 300 mmol L−1 acetic acid solution as the running buffer. Under the optimum conditions, the calibration curve showed good linearity (r 2 = 0.996), and the limit of detection was down to 0.12 ng mL−1 with the enrichment factor at 267. The relative standard deviation (RSD) values for peak area and migration time at a spiked concentration of 10 ng mL−1 of bromate were below 8.8 and 2.5 %, respectively. This proposed EME-CZE-C4D method has been successfully applied to analyze bottled drinking water and tap water samples with recoveries in the range of 85~98 %, providing an alternative to the determination of bromate in drinking water.