Monitoring of positron using high-energy gamma camera for proton therapy
详细信息    查看全文
  • 作者:Seiichi Yamamoto ; Toshiyuki Toshito ; Masataka Komori…
  • 关键词:Proton therapy ; Positron ; Gamma camera ; Imaging
  • 刊名:Annals of Nuclear Medicine
  • 出版年:2015
  • 出版时间:April 2015
  • 年:2015
  • 卷:29
  • 期:3
  • 页码:268-275
  • 全文大小:1,748 KB
  • 参考文献:1. Bennett, GW, Archambeau, JO, Archambeau, BE, Meltzer, JI, Wingate, CL (1978) Visualization and transport of positron emission from proton activation in vivo. Science 200: pp. 1151-1153 CrossRef
    2. Guerra, A, Domenico, G, Gambaccini, M, Marziani, M (1994) A Monte Carlo simulation of the possible use of positron emission tomography in proton radiotherapy. Nucl Instr Methods A 345: pp. 379-384 CrossRef
    3. Oelfke, U, Lam, GKY, Atkins, MS (1996) Proton dose monitoring with PET: quantitative studies in Lucite. Phys Med Biol 41: pp. 177-196 CrossRef
    4. Hishikawa, Y, Kagawa, K, Murakami, M, Sakai, H, Akagi, T, Abe, M (2002) Usefulness of positron-emission tomographic images after proton therapy. Int J Radiat Oncol Biol Phys 53: pp. 1388-1391 CrossRef
    5. Parodi, K, Paganetti, H, Shih, EA, Michaud, S, Loeffler, JS, Delaney, TF (2007) Patient study on in vivo verification of beam delivery and range using PET/CT imaging after proton therapy. Int J Radiat Oncol Biol Phys 68: pp. 920-934 CrossRef
    6. Litzenberg, DW, Roberts, DA, Lee, MY, Pham, K, Vander Molen, AM, Ronninge, R (1999) On-line monitoring of radiotherapy beams. Med Phys 26: pp. 992-1006 CrossRef
    7. Parodi, K, Enghardt, W, Haberer, T (2002) In-beam PET measurements of radioactivity induced by proton beams. Phys Med Biol 47: pp. 21-36 CrossRef
    8. Inaniwa, T, Tomitani, T, Kohno, T, Kanai, T (2005) Quantitative comparison of suitability of various beams for range monitoring with induced activity in hadron therapy. Phys Med Biol 50: pp. 1131-1145 CrossRef
    9. Iseki, Y, Mizuno, H, Futami, Y, Tomitani, T, Kanai, T, Kanazawa, M (2003) Positron camera for range verification of heavy-ion radiotherapy. Nucl Instrum Methods-A 515: pp. 840-849 CrossRef
    10. Nishio, T, Ogino, T, Nomura, K, Uchida, H (2006) Dose-volume delivery guided proton therapy using beam on-line PET system. Med Phys 33: pp. 4190-4197 CrossRef
    11. Yamaya, T, Yoshida, E, Inaniwa, T, Sato, S, Nakajima, Y, Wakizaka, H (2011) Development of a small prototype for a proof-of-concept of open PET imaging. Phys Med Biol 56: pp. 1123-1137 CrossRef
    12. Tashima, H, Yamaya, T, Yoshida, E, Kinouchi, S, Watanabe, M, Tanaka, E (2012) A single-ring open PET enabling PET imaging during radiotherapy. Phys Med Biol 57: pp. 4705-4718 CrossRef
    13. Smeets, J, Roellinghoff, F, Prieels, D, Stichelbaut, F, Benilov, A, Busca, P (2012) Prompt gamma imaging with a slit camera for real-time range control in proton therapy. Phys Med Biol 57: pp. 3371-3405 CrossRef
    14. Perali, I, Celani, A, Bombelli, L, Fiorini, C, Camera, F, Clementel, E (2014) Prompt gamma imaging of proton pencil beams at clinical dose rate. Phys Med Biol 59: pp. 5849-5871 CrossRef
    15. Kamada, K, Yanagida, T, Endo, T, Tsutsumi, K, Usuki, Y, Nikl, M (2012) 2 inch diameter single crystal growth and scintillation properties of Ce:Gd3Al2Ga3O12. J Cryst Growth 352: pp. 88-90 CrossRef
    16. Yamamoto, S, Watabe, H, Kato, K, Hatazawa, J (2012) Performance comparison of high quantum efficiency and normal quantum efficiency photomultiplier tubes and position sensitive photomultiplier tubes f
  • 刊物主题:Nuclear Medicine; Imaging / Radiology;
  • 出版者:Springer Japan
  • ISSN:1864-6433
文摘
Purpose In proton therapy, imaging of proton-induced positrons is a useful method to monitor the proton beam distribution after therapy. Usually, a positron emission tomography (PET) system installed in or near the proton beam treatment room is used for this purpose. However, a PET system is sometimes too large and expensive for this purpose. We developed a small field-of-view (FOV) gamma camera for high-energy gamma photons and used it for monitoring the proton-induced positron distribution. Methods The gamma camera used 0.85?mm?×?0.85?mm?×?10?mm Ce:Gd3Al2Ga3O12 (GAGG) pixels arranged in 20?×?20 matrix to form a scintillator block, which was optically coupled to a 1-inch-square position-sensitive photomultiplier tube (PSPMT). The GAGG detector was encased in a 20-mm thick container and a pinhole collimator was mounted on its front. The gamma camera was set 1.2?m from the 35?cm?×?35?cm?×?5?cm plastic phantom in the proton therapy treatment room, and proton beams were irradiated to the phantom with two proton energies. Results The gamma camera had spatial resolution of ~6.7?cm and sensitivity of 3.2?×?10? at 1?m from the collimator surface. For both proton energies, positron distribution in the phantom could be imaged by the gamma camera with 10-min acquisition. The lengths of the range of protons measured from the images were almost identical to the simulation results. Conclusions These results indicate that the developed high-energy gamma camera is useful for imaging positron distributions in proton therapy.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700