文摘
Many physics experiments depend on accurate high-voltage measurements to determine for example the exact retardation potential of an electron spectrometer as in the KATRIN experiment or the acceleration voltage of the ions at ISOL facilities. Until now only precision high-voltage dividers can be used to measure voltages up to 65 kV with an accuracy of 1 ppm. However, these dividers need frequent calibration and cross-checking and the direct traceability is not given. In this article we will describe the status of an experiment which aims to measure high voltages using collinear laser spectroscopy and which has the potential to provide a high-voltage standard and hence, a calibration source for precision high-voltage dividers on the 1 ppm level.