Calibration of CryojetHT and Cobra Plus Cryosystems used in X-ray diffraction studies
详细信息    查看全文
  • 作者:A. P. Dudka ; I. A. Verin ; E. S. Smirnova
  • 刊名:Crystallography Reports
  • 出版年:2016
  • 出版时间:July 2016
  • 年:2016
  • 卷:61
  • 期:4
  • 页码:692-696
  • 全文大小:353 KB
  • 刊物类别:Physics and Astronomy
  • 刊物主题:Physics
    Crystallography
    Russian Library of Science
  • 出版者:MAIK Nauka/Interperiodica distributed exclusively by Springer Science+Business Media LLC.
  • ISSN:1562-689X
  • 卷排序:61
文摘
CryoJetHT (Oxford Instruments) and Cobra Plus (Oxford Cryosystems) cryosystems, which are used for sample cooling in X-ray diffraction experiments, have been calibrated. It is shown that the real temperature in the vicinity of the sample differs significantly (the deviation is as high as 8–10 K at low temperatures) from the temperature recorded by authorized sensors of these systems. The calibration results are confirmed by measurements of the unit-cell parameters of GdFe3(BO3)4 single crystal in the temperature range of its phase transition. It is shown that, to determine the real temperature of a sample, one must perform an independent calibration of cryosystems rather than rely on their ratings.Original Russian Text © A.P. Dudka, I.A. Verin, E.S. Smirnova, 2016, published in Kristallografiya, 2016, Vol. 61, No. 4, pp. 663–668.