Syringe pump displacement alters line internal pressure and flow
详细信息    查看全文
  • 作者:Hiroshi Igarashi (1)
    Yukako Obata (1)
    Yoshiki Nakajima (1)
    Takasumi Katoh (1)
    Koji Morita (2)
    Shigehito Sato (1)
  • 刊名:Canadian Journal of Anesthesia/Journal canadien d'anesth篓娄sie
  • 出版年:2005
  • 出版时间:August 2005
  • 年:2005
  • 卷:52
  • 期:7
  • 页码:685-691
  • 全文大小:524KB
  • 参考文献:1. / Kern H, Kuring A, Redlich U, et al. Downward movement of syringe pumps reduces syringe output. Br J Anaesth 2001; 86: 828-1. CrossRef
    2. / Neff TA, Fischer JE, Schulz G, Baenziger O, Weiss M. Infusion pump performance with vertical displacement: effect of syringe pump and assembly type. Intensive Care Med 2001; 27: 287-1. CrossRef
    3. / Russel GB, Myers JL, Kofke WA. Care of the cardiac surgical patient: the first 24 hours postoperatively. / In: Hensley FA, Martin DE (Eds). The Practice of Cardiac Anesthesia. Boston: Little Brown; 1990: 289-21.
    4. / Sato S, Suzuki A, Nakajima Y, Iwamoto T, Bito H, Miyabe M. S-nitroso-N-acetylpenicillamine (SNAP) during hemorrhagic shock improves mortality as a result of recovery from vascular hyporeactivity. Anesth Analg 2000; 90: 362-. CrossRef
    5. / Weiss M, B?nziger O, Neff T, Fanconi S. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 2000; 26: 776-. CrossRef
    6. / Weiss M, Fischer J, Neff T, Baenziger O. The effects of syringe plunger design on drug delivery during vertical displacement of syringe pumps. Anaesthesia 2000; 55: 1094-. CrossRef
    7. / Weiss M, Hug MI, Neff T, Fischer J. Syringe size and flow rate affect drug delivery from syringe pumps. Can J Anesth 2000; 47: 1031-.
    8. / L?nnqvist PA. How continuous are continuous drug infusions? Intensive Care Med 2000; 26: 660-. CrossRef
    9. / L?nnqvist PA, L?fqvist B. Design flaw can convert commercially available continuous syringe pumps to intermittent bolus injectors. Intensive Care Med 1997; 23: 998-001. CrossRef
    10. / Klem SA, Farrington JM, Leff RD. Influence of infusion pump operation and flow rate on hemodynamic stability during epinephrine infusion. Crit Care Med 1993; 21: 1213-. CrossRef
    11. / Rooke GA, Bowdle TA. Syringe pumps for infusion of vasoactive drugs: mechanical idiosyncrasies and recommended operating procedures. Anesth Analg 1994; 78: 150-. CrossRef
    12. / Weiss M, Gerbe S, Füchslin RM, Neff TA. Accurate continuous drug delivery at low infusion rate with a novel microvolumetric infusion pump (MVIP): pump design, evaluation and comparison to the current standard. Anaesthesia, 2004; 59: 1133-. CrossRef
    13. / Neff TA, Fellmann C, Fuechslin RM, Gerber AC, Weiss M. The Panomat? P-10 micro-volumetric infusion pump is suitable for continuous drug administration at minimal flow rates. Can J Anesth 2002; 49: 1048-2. CrossRef
  • 作者单位:Hiroshi Igarashi (1)
    Yukako Obata (1)
    Yoshiki Nakajima (1)
    Takasumi Katoh (1)
    Koji Morita (2)
    Shigehito Sato (1)

    1. Anesthesiology and Intensive Care, Hamamatsu University School of Medicine, 1-20-1 Handayama, 431-3192, Hamamatsu, Japan
    2. the Surgical Center, Hamamatsu University School of Medicine, Hamamatsu, Japan
文摘
Purpose It has been reported that the actual flow from a syringe pump changes due to vertical movement of the pumpin vitro, but a direct study of thein vivo effects of fluid delivery irregularities caused by vertical pump displacement has not been performed. The aim of this study was to assess the influence of positional changes of the syringe pump on the internal pressure (IP) and flow from the circuit, and to examine blood pressure changes caused by pump movement in animals with or without hemorrhagic shock. Methods To simulate clinical conditions, we used a rabbit model. We first measured the changes in line IP and flow from the syringe pump after moving the pump 50 cm vertically upwards or downwards. With the same animal, we then recorded the blood pressure changes under these conditions during norepinephrine (NE) infusion, using a rabbit hemorrhagic shock (HS) model. Results Following downward movement of the syringe pump, the IP increased by 37.5 ± 4.0 mmHg and IP decreased by 37.3 ± 3.1 mmHg following upward movement of the syringe pump. In the rabbit HS model, movement downwards decreased systolic blood pressure (SBP) by -17 ± 6.9% (P < 0.001), while upward movement raised SBP by 45.7 ± 21.5% (P < 0.001) from baseline values. Conclusions: Vertical displacement of the syringe pump alters the flow due to a change of line I P, and blood pressure can be affected by pump movement during NE infusion.