Improved MRI of the neonatal heart: feasibility study using a knee coil
详细信息    查看全文
  • 作者:Michael Helle (13)
    Michael Jerosch-Herold (2)
    Inga Voges (1)
    Chris Hart (1)
    Hans-Heiner Kramer (1)
    Carsten Rickers (1) rickers@pedcard.uni-kiel.de
  • 关键词:Cardiac magnetic resonance imaging &#8211 ; Neonate &#8211 ; Transmit and receive MRI extremity coil &#8211 ; 3 ; Tesla MRI
  • 刊名:Pediatric Radiology
  • 出版年:2011
  • 出版时间:November 2011
  • 年:2011
  • 卷:41
  • 期:11
  • 页码:1429-1432
  • 全文大小:105.5 KB
  • 参考文献:1. Pennell DJ, Sechtem UP, Higgins CB et al (2004) Clinical indications for cardiovascular magnetic resonance (CMR): Consensus Panel report. Eur Heart J 25:1940–1965
    2. Groves AM (2011) Cardiac magnetic resonance in the study of neonatal haemodynamics. Semin Fetal Neonatal Med 16:36–41
    3. Duerden RM, Pointon KS, Habib S (2006) Review of clinical cardiac MRI. Imaging 18:178–186
    4. Gutberlet M, Spors B, Grothoff M et al (2004) Comparison of different cardiac MRI sequences at 1.5/3.0 T with respect to signal-to-noise and contrast-to-noise ratios: initial experience. R枚fo 176:801–808
    5. Pamboucas C, Nihoyannopoulos P (2006) Cardiovascular magnetic resonance at 3 Tesla: advantages, limitations and clinical potential. Hellenic J Cardiol 47:170–173
    6. Schick F (2005) Whole-body MRI at high field: technical limits and clinical potential. Eur Radiol 15:946–959
    7. Wen H, Denison TJ, Singeman RW et al (1997) The intrinsic signal-to-noise ratio in human cardiac imaging at 1.5, 3 and 4 T. J Magn Reson 125:65–71
    8. Wieben O, Francois C, Reeder SB (2008) Cardiac MRI of ischemic heart disease at 3 T: potential and challenges. Eur J Radiol 65:15–28
    9. Groves AM, Chiesa G, Durighel G et al (2011) Functional cardiac MRI in preterm and term newborns. Arch Dis Child Fetal Neonatal Ed 96:86–91
    10. Kawel N, Valsangiacomo-Buechel E, Hoop R et al (2010) Preoperative evaluation of pulmonary artery mophology and pulmonary circulation in neonates with pulmonary atresia—usefulness of MR angiography in clinical routine. J Cardiovasc Magn Reson 12:52–59
    11. Merchant N, Groves A, Larkman DJ et al (2009) A patient care system for early 3.0 Tesla magnetic resonance imaging of very low birth weight infants. Early Hum Dev 85:779–783
  • 作者单位:1. Department of Pediatric Cardiology, Christian-Albrechts-Universit盲t, Universit盲tsklinikum Schleswig-Holstein, Campus Kiel, Arnold-Heller-Str. 3, 24105 Kiel, Germany2. Brigham and Women鈥檚 Hospital, Harvard Medical School, Boston, USA3. Institute of Neuroradiology, Christian-Albrechts-Universit盲t, Universit盲tsklinikum Schleswig-Holstein, Campus Kiel, Kiel, Germany
  • 刊物类别:Medicine
  • 刊物主题:Medicine & Public Health
    Imaging and Radiology
    Pediatrics
    Neuroradiology
    Nuclear Medicine
    Ultrasound
    Oncology
  • 出版者:Springer Berlin / Heidelberg
  • ISSN:1432-1998
文摘
As a first step towards the development of a neonatal cardiac MRI (CMRI) coil, we investigated the utilization, handling and image quality of a knee coil for CMRI in infants at 3 Tesla. Scout imaging and cine imaging with optimized pulse sequence parameters were performed in two neonates and the resulting images were evaluated for diagnostic purposes. All images were of high diagnostic quality. Complete examinations took less than five minutes. The excellent signal-to-noise ratio provided by this coil allowed all cine studies to be acquired with only three averages. In conclusion, a knee coil can be successfully applied in neonates for CMRI, but the coil design could be improved further to facilitate routine clinical use.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700