Clinical evaluation of the polymethylmethacrylate-augmented thoracic and lumbar pedicle screw fixation guided by the three-dimensional navigation for the osteoporosis patients
详细信息    查看全文
  • 作者:Qiang Yuan ; Guilin Zhang ; Jingye Wu ; Yonggang Xing ; Yuqing Sun
  • 关键词:PMMA ; Augmentation ; Navigation guided ; Cement leakage ; Osteoporosis ; Pedicle screw fixation
  • 刊名:European Spine Journal
  • 出版年:2015
  • 出版时间:May 2015
  • 年:2015
  • 卷:24
  • 期:5
  • 页码:1043-1050
  • 全文大小:2,081 KB
  • 参考文献:1.Cummings SR, Melton LJ (2002) Epidemiology and outcomes of osteoporotic fractures. Lancet 359(9319):1761-767View Article PubMed
    2.Paxinos O, Tsitsopoulos PP, Zindrick MR, Voronov LI et al (2010) Evaluation of pullout strength and failure mechanism of posterior instrumentation in normal and osteopenic thoracic vertebrae. J Neurosurg Spine 13(4):469-76View Article PubMed
    3.Halvorson TL, Kelley LA, Thomas KA, Whitecloud TS 3rd, Cook SD (1994) Effects of bone mineral density on pedicle screw fixation. Spine (Phila Pa 1976) 19(21):2415-420View Article
    4.Garfin SR, Yuan HA, Reiley MA (2001) New technologies in spine: kyphoplasty and vertebroplasty for the treatment of painful osteoporotic compression fractures. Spine (Phila Pa 1976) 26(14):1511-515View Article
    5.Frankel BM, Jones T, Wang C (2007) Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery 61(3):531-37 (discussion 537-38)View Article PubMed
    6.Fuentes S, Blondel B, Metellus P, Gaudart J, Adetchessi T, Dufour H (2010) Percutaneous kyphoplasty and pedicle screw fixation for the management of thoraco-lumbar burst fractures. Eur Spine J 19(8):1281-287View Article PubMed Central PubMed
    7.Becker S, Chavanne A, Spitaler R, Kropik K et al (2008) Assessment of different screw augmentation techniques and screw designs in osteoporotic spines. Eur Spine J 17(11):1462-469View Article PubMed Central PubMed
    8.Gao M, Lei W, Wu Z, Liu D, Shi L (2011) Biomechanical evaluation of fixation strength of conventional and expansive pedicle screws with or without calcium based cement augmentation. Clin Biomech (Bristol, Avon) 26(3):238-44View Article
    9.Waits C, Burton D, McIff T (2009) Cement augmentation of pedicle screw fixation using novel cannulated cement insertion device. Spine (Phila Pa 1976) 34(14):E478–E483View Article
    10.Ren H, Shen Y, Zhang YZ, Ding WY et al (2010) Correlative factor analysis on the complications resulting from cement leakage after percutaneous kyphoplasty in the treatment of osteoporotic vertebral compression fracture. J Spinal Disord Tech 23(7):e9–e15View Article PubMed
    11.Wilkes RA, Mackinnon JG, Thomas WG (1994) Neurological deterioration after cement injection into a vertebral body. J Bone Joint Surg Br 76(1):155PubMed
    12.Rajan VV, Kamath V, Shetty AP, Rajasekaran S (2010) Iso-C3D navigation assisted pedicle screw placement in deformities of the cervical and thoracic spine. Indian J Orthop 44 (2):163-68
    13.Kosmopoulos V, Schizas C (2007) Pedicle screw placement accuracy: a meta-analysis. Spine (Phila Pa 1976) 32(3):E111–E120View Article
    14.Engelke K, Adams JE, Armbrecht G, Augat P et al (2008) Clinical use of quantitative computed tomography and peripheral quantitative computed tomography in the management of osteoporosis in adults: the 2007 ISCD official positions. J Clin Densitom 11(1):123-62View Article PubMed
    15.Sawakami K, Yamazaki A, Ishikawa S, Ito T, Watanabe K, Endo N (2012) Polymethylmethacrylate augmentation of pedicle screws increases the initial fixation in osteoporotic spine patients. J Spinal Disord Tech 25(2):E28–E35View Article PubMed
    16.Gaines RW Jr (2000) The use of pedicle-screw internal fixation for the operative treatment of spinal disorders. J Bone Joint Surg Am 82-A(10):1458-476PubMed
    17.Lane NE (2006) Epidemiology, etiology, and diagnosis of osteoporosis. Am J Obstet Gynecol 194(2 Suppl):S3–S11View Article PubMed
    18.Sarzier JS, Evans AJ, Cahill DW (2002) Increased pedicle screw pullout strength with vertebroplasty augmentation in osteoporotic spines. J Neurosurg 96(3 Suppl):309-12PubMed
    19.Wittenberg RH, Lee KS, Shea M, White AA 3rd, Hayes WC (1993) Effect of screw diameter, insertion technique, and bone cement augmentation of pedicular screw fixation strength. Clin Orthop Relat Res 296:278-87PubMed
    20.Frankel BM, Jones T, Wang C (2007) Segmental polymethylmethacrylate-augmented pedicle screw fixation in patients with bone softening caused by osteoporosis and metastatic tumor involvement: a clinical evaluation. Neurosurgery 61(3):531-37 (discussion 537-38)View Article PubMed
    21.Chang MC, Liu CL, Chen TH (2008) Polymethylmethacrylate augmentation of pedicle screw for osteoporotic spinal surgery: a novel technique. Spine (Phila Pa 1976) 33(10):E317–E324View Article
    22.Rajasekaran S, Vidyadhara S, Ramesh P, Shetty AP (2007) Randomized clinical study to compare the accuracy of navigated and non-navigated thoracic pedicle screws in deformity correction surgeries. Spine (Phila Pa 1976) 32(2):E56–E64View Article
    23.Hulme PA, Krebs J, Ferguson SJ, Berlemann U (2006) Vertebroplasty and kyphoplasty: a systematic review of 69 clinical studies. Spine (Phila Pa 1976) 31(17):1983-001View Article
    24.Merloz P, Troccaz J, Vouaillat H, Vasile C et al (2007) Fluoroscopy-based navigation system in s
  • 作者单位:Qiang Yuan (1)
    Guilin Zhang (1)
    Jingye Wu (1)
    Yonggang Xing (1)
    Yuqing Sun (1)
    Wei Tian (1)

    1. Department of Spine Surgery, Beijing Jishuitan Hospital, Peking University Fourth Clinical Medical College, 31 Xinjiekou East Street, Xicheng District, Beijing, 100035, China
  • 刊物类别:Medicine
  • 刊物主题:Medicine & Public Health
    Surgical Orthopedics
    Neurosurgery
  • 出版者:Springer Berlin / Heidelberg
  • ISSN:1432-0932
文摘
Purpose To evaluate the safety and efficacy of three-dimensional (3D) navigation-guided polymethylmethacrylate (PMMA)-augmented thoracic and lumbar pedicle screw fixation for the osteoporotic patients. Methods 27 consecutive osteoporosis patients with a variety of spinal disorders who underwent 3D navigation-guided PMMA-augmented pedicle screw fixation were evaluated clinically and radiologically in the perioperative and 1-year follow-up period. The improvement of Japanese Orthopaedic Association (JOA) scores was analyzed. PMMA leakage and other complications were inspected intraoperatively and postoperatively. Screw loosening and bone fusion were evaluated radiographically during follow-up. Results 8 patients had thoracic and lumbar fractures; 18 patients had degenerative spinal disorders; one patient had revision surgery. One patient died of postoperative pneumonia. Each of the other 26 patients was followed up regularly at 3, 6, 12 and 18?months postoperatively. The mean therapeutic improvement rate is 39.6?% evaluated by JOA scores. 2 patients (7.4?%) had leakage into the spinal canal in front of the posterior longitudinal ligament and two patients (7.4?%) had leakage into the prevertebral soft tissue inspected by the postoperative CT scans. No pedicle cortex breach and cement leakage surrounding pedicle cortex were observed. None of patients complained of dyspnoea and showed evidence of pulmonary embolism. Bone fusions were found in 20 patients (bony fusion rate 76.9?%) at the 12-month follow-up and no screw loosening occurs. Conclusion The results show favorable outcome using 3D navigation-guided PMMA-augmented thoracic and lumbar pedicle screw fixation for the osteoporosis patients both clinically and radiologically.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700