Determination of vanadium in titanate-based ferroelectrics by INAA with discriminating gamma-ray spectrometry
详细信息    查看全文
文摘
Instrumental neutron activation analysis was used for determination of vanadium mass fraction in crystals of incipient ferroelectric strontium titanate and ferroelectric barium titanate. In order to improve vanadium limit of detection, discriminating gamma-ray spectrometry was used by inserting an absorption filter between the samples and an HPGe detector. The use of the absorption lead filter 6-mm thick yielded improvement of the vanadium limit of detection by a factor of two. The vanadium mass fraction determined in a quality control sample, which was NIST standard reference material SRM 1648 Urban Particulate Matter, was in close agreement with the certified value.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700