Effects of refractive blur on the multifocal electroretinogram
详细信息    查看全文
文摘
A significant difference in the response density of the MF-ERG response has been suggested for every 2 diopter change of refraction. The influence of refractive blur on the MF-ERG was studied in 8 healthy volunteers using either the VERIS™ system (Group A: n=5) or Retiscan™ (Group B: n=3). For each eye recordings were obtained with a corrective lens of −3 dpt, 0 dpt, +3 dpt and +6 dpt placed in front of the contact lens electrode. The viewing distance was adjusted to compensate for the induced changes in the retinal image size. When the changes in retinal image size due to the refractive lens were compensated for, no influence due to refraction was observed in either latencies or amplitudes of (KI (P>0.05). This held true for the central response average (four degrees) as well as for the outer 6–25 degrees. In KII.1 only the peripheral amplitudes of Group B showed an influence due to refraction (P≤0.05). This may be due to adaptation as the recordings of group B were obtained in succession. As expected, significant differences were observed when the recordings obtained with the different systems were compared (P≤0.05).

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700