Nano-Au/cMWCNT Modified speB Gene Specific Amperometric Sensor for Rapidly Detecting Streptococcus pyogenes causing Rheumatic Heart Disease
详细信息    查看全文
文摘
A specific 5′ NH2 labeled DNA probe of speB gene was immobilized onto the gold nanoparticles/carboxylated multi walled carbon nanotubes (Nano-Au/cMWCNT) screen printed electrode using EDC/NHS cross linking chemistry. This was followed by hybridization with 0.5–50 ng/6 µl of single stranded genomic DNA Streptococcus pyogenes infected patient throat swab samples. Electrochemical amperometric assay was deciphered by using cyclic voltammetry (CV) with methylene blue a redox indicator. The sensor had a sensitivity of 104.7 µA cm−2 ng−1 using CV with a R2 of 0.907 and 0.01 ng/6 µl as the limit of detection (LOD). The modified electrode surface morphology was characterized using scanning electron microscopy. The stability of the electrode was seen at 4 °C for 180 days having 6% loss in the initial current. The sensor is speB gene specific and can detect the pathogen within 30 min.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083