Application of ICP-QMS for the determination of plutonium in environmental samples for safeguards purposes
详细信息查看全文 | 推荐本文 |
摘要
Aiming to determine the plutonium amount as well as its isotopic composition, in particular, in swipe samples for safeguards purposes, an analytical method was developed with a plutonium separation step based on extraction chromatography using 2 cm TEVA columns and detection with quadrupole ICP-MS applying an ultra-sonic nebulizer coupled with membrane desolvation system. The method was successfully applied to New Brunswick plutonium certified reference materials as well as to Lawrence Livermore National Laboratory round robin samples, based on the round robin samples provided by the Institute for Reference Materials and Measurements (Belgium), as part of the Regular European Interlaboratory Measurement Evaluation Programme (REIMEP), campaign 16 (isotopic abundances of plutonium in plutonium nitrate samples), with a total plutonium amount between 1 and 0.25 ng per sample. After the introduction of an additional separation step, it was also possible to carry out precise and accurate total plutonium, p>240p>Pu/p>239p>Pu, p>241p>Pu/p>239p>Pu and p>242p>Pu/p>239p>Pu atom ratios determination in sediment sample showing its applicability to environmental samples in general, reaching a detection limit equivalent to 5 mBq p>239p>Pu kgp>−1p>.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700