One hundred and nineteen patients with 277 grafts were randomized to Group 1 based on BMI-adapted scanning protocol with prospective ECG-triggering (40 patients), Group 2 with prospective ECG-triggering (39 patients) and Group 3 (40 patients) with retrospective ECG-triggering. Data were acquired using 64-slice MDCT.
MDCT correctly assessed the patency of all CABG in 3 groups. After comparison with ICA, MDCT was able to correctly detect the occlusion or stenosis of CABG in all groups, with the exception of one case of Group 3. In Group 3 sensitivity, specificity, positive predictive value, negative predictive value and accuracy of CABG evaluation were 100%, 98.4%, 96.7%, 100%and 98.9%, respectively. In Groups 1 and 2 the diagnostic accuracy of CABG evaluation was 100%. Effective radiation dose was 3.5 卤 1.4 mSv in Group 1 vs. 7.4 卤 2.6 mSv in Group 2 vs. 27.8 卤 9.4 mSv in Group 3.
Our results indicated that MDCT-XTe and MDCT-XT have a diagnostic performance in the evaluation of CABG similar to MDCT-Helical, with a significant reduction of radiation exposure, specially for MDCT-XTe.