Potentiometric enzyme electrode for urea determination using immobilized urease in poly(vinylferrocenium) film
详细信息查看全文 | 推荐本文 |
摘要
A new enzyme electrode for the determination of urea was developed by immobilizing urease in poly(vinylferrocenium) (PVF+) matrix. A PVF+ClO4 film was coated on Pt electrode at +0.7 V by electrooxidation of poly(vinylferrocene) in metylene chloride containing 0.1 M tetrabutylammonium perchlorate (TBAP). The enzyme modified electrode PVF+E was prepared by anion-exchange in an enzyme solution in 1.0 mM phosphate buffer at pH 7.0. FTIR spectroscopy was used to identify PVF+ClO4, and PVF+E. UV spectroscopy was also used to prove the enzyme immobilization. The effects of polymeric film thickness, concentration of enzyme solution, immobilization time of the enzyme, pH and temperature of the medium, concentration of the buffer solution and possible interferents on the measured potential values were investigated. The potentiometric enzyme electrode developed in this study provided linearity to urea in the 5 × 10−5 to 1 × 10−1 M urea concentration range. The detection limit under the optimum working conditions was determined as 5 × 10−6 M for urea. The enzyme electrode was found to be stable for 24 days. The apparent Michaelis-Menten consant (KM app) value and the activation energy, Ea, of this immobilized enzyme system were found to be 4.48 × 10−5 M and 4.97 kcal mol−1 for urea, respectively.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700