Comparison of localization performance with implanted fiducial markers and cone-beam computed tomography for on-line image-guided radiotherapy of the prostate
详细信息查看全文 | 推荐本文 |
摘要
Purpose: The aim of this work was to assess the accuracy of kilovoltage (kV) cone-beam computed tomography (CBCT)–based setup corrections as compared with orthogonal megavoltage (MV) portal image-based corrections for patients undergoing external-beam radiotherapy of the prostate.

Methods and Materials: Daily cone-beam CT volumetric images were acquired after setup for patients with three intraprostatic fiducial markers. The estimated couch shifts were compared retrospectively to patient adjustments based on two orthogonal MV portal images (the current clinical standard of care in our institution). The CBCT soft-tissue based shifts were also estimated by digitally removing the gold markers in each projection to suppress the artifacts in the reconstructed volumes. A total of 256 volumetric images for 15 patients were analyzed.

Results: The Pearson coefficient of correlation for the patient position shifts using fiducial markers in MV vs. kV was (R2 = 0.95, 0.84, 0.81) in the left–right (LR), anterior–posterior (AP), and superior–inferior (SI) directions, respectively. The correlation using soft-tissue matching was as follows: R2 = 0.90, 0.49, 0.51 in the LR, AP and SI directions. A Bland-Altman analysis showed no significant trends in the data. The percentage of shifts within a ±3-mm tolerance (the clinical action level) was 99.7%, 95.5%, 91.3%for fiducial marker matching and 99.5%, 70.3%, 78.4%for soft-tissue matching.

Conclusions: Cone-beam CT is an accurate and precise tool for image guidance. It provides an equivalent means of patient setup correction for prostate patients with implanted gold fiducial markers. Use of the additional information provided by the visualization of soft-tissue structures is an active area of research.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700