The sensitivity and specificity of ultrasound estimation of central venous pressure using the internal jugular vein
详细信息查看全文 | 推荐本文 |
摘要

Purpose

The fluid volume status of a patient is difficult to assess clinically. The aim of this study was to compare the ultrasound estimation of the height of the right internal jugular vein (CVPIJV) with direct estimation of central venous pressure (CVP) (CVPCVC).

Materials and Methods

A portable ultrasound machine defined the 鈥渢op鈥?of the right internal jugular vein in 44 patients from a single tertiary hospital. The vertical height from this point to the sternal angle was used to estimate CVPIJV. A central venous catheter was then inserted and direct measurement of CVP was made with a pressure transducer. A normal CVP was defined as 3 to 6 mm Hg.

Results

For overloaded patients, CVPIJV correlated well with CVPCVC, P = .004, sensitivity of 64.3%, specificity of 81.3%, and positive predictive value of 85.7%. The area under the curve for the receiver operating characteristic curve was 0.73 (95%confidence interval, 0.59-0.86). For undervolumed patients, the correlation remained statistically significant, P < .001, sensitivity of 88.9%, specificity of 77.1%, and negative predictive value of 96.4%. The area under the curve was 0.83 (95%confidence interval, 0.70-0.96).

Conclusion

Ultrasound estimation of CVP using a portable ultrasound machine and the internal jugular vein is simple, noninvasive, and accurate.