Simultaneous electrokinetic and hydrodynamic injection with on-line sample concentration via micelle to solvent stacking in micellar electrokinetic chromatography
详细信息查看全文 | 推荐本文 |
摘要
Simultaneous electrokinetic and hydrodynamic injection (SEHI) of organic cations (tricyclic antidepressant and beta blocker drugs) with on-line sample concentration using micelle to solvent stacking (MSS) was studied in micellar electrokinetic chromatography. Compared to conventional injection, >300-fold improvements in signals were obtained by hydrodynamic injection. However, with SEHI the amount of sample ions introduced into the capillary was increased which afforded a higher gain of up to 4000-fold without compromise to separation efficiency. The electrokinetic injection at negative polarity (anode at the detector end) introduced the micelle bound analytes. The hydrodynamic injection also maintained the MSS boundary inside the capillary. The stability of the MSS boundary affected SEHI where mild conditions that were low voltage as well as pressure injection were desired. The limits of detection were in the range from 0.6-4.2 ng mL鈭?. A strategy for optimization was described and the method was applied to the ng mL鈭? analysis of spiked wastewater after simple dilution and centrifugation.