Determination of bisphenol A in river water and body fluid samples by stir bar sorptive extraction with in situ derivatization and thermal desorption-gas chromatography–mass spectrometry
详细信息查看全文 | 推荐本文 |
摘要
A new method, based on stir bar sorptive extraction (SBSE) with in situ derivatization and thermal desorption (TD)-gas chromatography&ndash;mass spectrometry (GC&ndash;MS) is described for the determination of trace amounts of bisphenol A (BPA) in river water, urine, plasma, and saliva samples. The derivatization conditions with acetic acid anhydride and the SBSE conditions such as sample volumes and extraction time are investigated. Then, the stir bar is subjected to TD followed by GC&ndash;MS. The detection limits of BPA in river water, urine, plasma, and saliva samples are 1&ndash;5, 20, 100, and 20pgml−1 (ppt), respectively. Calibration for BPA was shown to be linear with a correlation coefficient of >0.99. The average recoveries of BPA in all samples are higher than 95%(R.S.D. < 10%) with correction using an added surrogate standard, 13C12-bisphenol A. This simple, accurate, sensitive, and selective analytical method may be applicable to the determination of trace amounts of BPA in liquid samples.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700