Prompt gamma tests of LaBr3:Ce and BGO detectors for detection of hydrogen, carbon and oxygen in bulk samples
详细信息查看全文 | 推荐本文 |
摘要
Prompt gamma ray tests of cylindrical lanthanum halide (LaBr3:Ce) and bismuth germanate (BGO) gamma ray detectors have been carried out for detection of hydrogen, carbon and oxygen concentrations in bulk samples via inelastic scattering of neutrons using a 14 MeV neutron-based prompt gamma neutron activation analysis setup. Regardless of its intrinsic activity, the LaBr3:Ce detector showed superior performance than the BGO detector for the detection of hydrogen, carbon and oxygen concentrations in benzene, water, toluene, propanol, ethanol and methanol bulk samples. The BGO detector has a large concentration of oxygen in its detector material and is consequently less sensitive for oxygen detection in bulk samples. Hence, it is not a suitable choice for oxygen determination in bulk samples.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700