摘要
Prompt gamma ray tests of cylindrical lanthanum halide (LaBr3:Ce) and bismuth germanate (BGO) gamma ray detectors have been carried out for detection of hydrogen, carbon and oxygen concentrations in bulk samples via inelastic scattering of neutrons using a 14 MeV neutron-based prompt gamma neutron activation analysis setup. Regardless of its intrinsic activity, the LaBr3:Ce detector showed superior performance than the BGO detector for the detection of hydrogen, carbon and oxygen concentrations in benzene, water, toluene, propanol, ethanol and methanol bulk samples. The BGO detector has a large concentration of oxygen in its detector material and is consequently less sensitive for oxygen detection in bulk samples. Hence, it is not a suitable choice for oxygen determination in bulk samples.