摘要
Electro-osmotic characteristics such as the electro-osmotic velocity, the electro-osmotic mobility and zeta-potential were determined in polystyrene microfluidic chips by the conductivity (electric current monitoring) method. The capillary microchips were fabricated with the use of micromilling and hot assembling techniques. A linear dependence of the electro-osmotic velocity on the voltage applied on 39 mm long microcapillaries was obtained. The electro-osmotic mobility and zeta-potential were found to be equal to 2.26 ± 0.12 × 10−8 m2 V−1 s−1 and −29.0 ± 1.5 mV, respectively.