End-column chemiluminescence detection for pressurized capillary electrochromatographic analysis of norepinephrine and epinephrine
详细信息查看全文 | 推荐本文 |
摘要
A simple and convenient end-column chemiluminescence (CL) detection coupled to pressurized capillary electrochromatography (pCEC) was described. Luminol and N-(4-aminobutyl)-N-ethylisoluminol (ABEI) were adopted as mode compounds to evaluate the feasibility of end-column reactor. Detailed analysis of ABEI revealed that the high sensitivity could be obtained with the reactor. Furthermore, determination of norepinephrine (NE) and epinephrine (EP), which were labeled with ABEI, was accomplished by using the end-column pCEC–CL detection based on ABEI–potassium ferricyanide–alkaline medium CL reaction system. Under the optimum conditions, the detection limit (S/N = 3) of NE and EP was 0.08 μM and 0.06 μM, respectively. The proposed method has also been successfully applied to the analysis of adrenaline hydrochloride injection sample.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700