A simple and reusable fluorescent sensor for heme proteins based on a conjugated polymer-doped electrospun nanofibrous membrane
详细信息查看全文 | 推荐本文 |
摘要
We reported a simple and reusable fluorescent sensor for heme proteins based on the electrospun nanofibrous membrane doped with a fluorescent conjugated polymer P. The sensor showed favorable fluorescence sensing performance towards the heme proteins, including hemoglobin (Hb), myoglobin (Mb) and cytochrome c (Cyt c). The surface wettability and sensing performance of the electrospun nanofibrous membrane were investigated in detail using Hb as the model. The nanofibrous sensor showed satisfactory reversibility with less than 10%signal loss after nine quenching-regeneration cycles, and good batch-to-batch reproducibility with a relative standard deviation of 3.4%(n = 3). The linear range of the sensor for Hb determination was 2.0 脳 10鈭? to 3.0 脳 10鈭? M with a detection limit of 1.2 脳 10鈭? M. The quenching process is mainly based on the fluorescence resonance energy transfer mechanism between the fluorescent conjugated polymer P and the heme prosthetic groups, therefore the sensor was selective against most of the common interferents. As an example to evaluate the feasibility of the sensor in practical application, Hb in human blood samples was determined and the results were in good agreement with the data provided by the hospital. To the best of our knowledge, this is the first work using fluorescent electrospun nanofibrous sensor for protein analysis in real biological sample.

© 2004-2018 中国地质图书馆版权所有 京ICP备05064691号 京公网安备11010802017129号

地址:北京市海淀区学院路29号 邮编:100083

电话:办公室:(+86 10)66554848;文献借阅、咨询服务、科技查新:66554700