An ORS-ICP-MS method for monitoring trace levels of cobalt and chromium in whole blood samples from hip arthroplasty patients with metal-on-metal prostheses
详细信息查看全文 | 推荐本文 |
摘要

Objective

To develop a rapid and reliable method, using an octopole reaction system (ORS) ICP-MS, capable of monitoring trace levels of Co and Cr in whole blood samples from hip arthroplasty patients with metal-on-metal prostheses.

Design and method

Whole blood is diluted 10-fold with an alkaline diluent and analyzed using an Agilent 7500 CE ORS-ICP-MS.

Results

Limit of quantification of 0.03 渭g/L Co and 0.20 渭g/L Cr in patient samples. < 6%covariance obtained for quality control materials analyzed over 10 runs.

Conclusion

This method is capable of monitoring trace levels of Co and Cr in diluted whole blood samples with a vial to vial run time of approximately 2 min. Results are comparable to those obtained using high resolution (HR) ICP-MS with sample digestion.